Instrument Parameters
DART-TOFMS Instrumental Parameters
DART-TOFMS Parameter |
Conditions |
Ionization Gas |
Helium |
Gas Temperature |
350°C |
Orifice 1 Voltage |
20V |
Orifice 2 Voltage |
5V |
Ring Lens Voltage |
5V |
Rf Ion Guide Voltage |
600V |
m/z Range |
60-1000 amu |
Ionization Modes |
Positive and Negative |
GC-MS Instrumental Parameters
GC Parameter |
Conditions |
Column |
DB-5ms Ultra Inert |
Injection Volume |
1 μL |
Split Ratio |
100:1 |
Injection Temperature |
250°C |
Oven Initial Temperature (Hold Time) |
60°C (2 min) |
Oven Temperature Ramp Rate |
10°C /min |
Oven Temperature (Hold Time) |
320°C (15 min) |
Column Flow Rate |
1.2 mL/min |
MS Source Temperature |
230°C |
MS Quad Temperature |
150°C |
m/z Range |
30-550 amu |
ATR-FTIR Instrumental Parameters
FTIR Parameters |
Conditions |
Beam Splitter |
Ge/KBr |
Detector |
DLaTGS |
Range |
4000-400 cm-1 |
Scans |
60 |
Resolution |