INSTRUMENT CONTROL PARAMETERS: MSD-2 --------------------------------------- C:\MSDChem\1\METHODS\SP.m Control Information ------- ----------- Sample Inlet : GC Injection Source : Manual Injection Location : Front Use MS : Yes ============================================================================= 6890 GC METHOD ============================================================================= OVEN Initial temp: 45 'C (On) Maximum temp: 325 'C Initial time: 3.00 min Equilibration time: 0.50 min Ramps: # Rate Final temp Final time 1 15.00 150 0.00 2 40.00 265 7.00 3 0.0(Off) Post temp: 0 'C Post time: 0.00 min Run time: 19.88 min FRONT INLET (SPLIT/SPLITLESS) BACK INLET (SPLIT/SPLITLESS) Mode: Split Mode: Split Initial temp: 170 'C (On) Initial temp: 250 'C (Off) Pressure: 8.23 psi (On) Pressure: 0.00 psi (Off) Split ratio: 10:1 Total flow: 45.0 mL/min Split flow: 21.3 mL/min Gas saver: Off Total flow: 26.4 mL/min Gas type: Helium Gas saver: On Saver flow: 20.0 mL/min Saver time: 25.00 min Gas type: Helium COLUMN 1 COLUMN 2 Capillary Column Capillary Column Model Number: J & W 1225532 Model Number: Agilent 19091J-233 DB5 MS HP-5 5% Phenyl Methyl Siloxane Max temperature: 350 'C Max temperature: 325 'C Nominal length: 30.0 m Nominal length: 30.0 m Nominal diameter: 250.00 um Nominal diameter: 250.00 um Nominal film thickness: 0.25 um Nominal film thickness: 1.00 um Mode: constant flow Mode: (see column 1) Initial flow: 1.2 mL/min Nominal initial flow: 1.1 mL/min Nominal init pressure: 9.14 psi Nominal init pressure: 9.13 psi Average velocity: 39 cm/sec Average velocity: 39 cm/sec Inlet: Front Inlet Inlet: Front Inlet Outlet: MSD Outlet: MSD Outlet pressure: vacuum Outlet pressure: vacuum FRONT DETECTOR (NO DET) BACK DETECTOR (NO DET) SIGNAL 1 SIGNAL 2 Data rate: 20 Hz Data rate: 20 Hz Type: test plot Type: test plot Save Data: Off Save Data: Off Zero: 0.0 (Off) Zero: 0.0 (Off) Range: 0 Range: 0 Fast Peaks: Off Fast Peaks: Off Attenuation: 0 Attenuation: 0 COLUMN COMP 1 COLUMN COMP 2 (No Detectors Installed) (No Detectors Installed) THERMAL AUX 1 AUX PRESSURE 3 Use: MSD Transfer Line Heater Description: Description: Gas Type: Helium Initial temp: 280 'C (On) Initial pressure: 0.00 psi (Off) Initial time: 0.00 min # Rate Final temp Final time 1 0.0(Off) AUX PRESSURE 4 AUX PRESSURE 5 Description: Description: Gas Type: Helium Gas Type: Helium Initial pressure: 0.00 psi (Off) Initial pressure: 0.00 psi (Off) POST RUN Post Time: 0.00 min TIME TABLE Time Specifier Parameter & Setpoint GC Injector Front Injector: No parameters specified Back Injector: No parameters specified Column 1 Inventory Number : AB006 Column 2 Inventory Number : AB004 GERSTEL MPS Liquid Injection Syringe : 10ul SAMPLE PARAMETERS Inj. Volume : 1.0 uL Inj. Speed : 100.00 uL/s Fill Volume : 5.0 uL Fill Strokes : 2 Fill Speed : 5.00 uL/s Eject Speed : 100.00 uL/s Viscositiy Delay : 2.0 s Air Volume : 0.0 uL Pre Inj. Delay : 0.00 s Post Inj. Delay : 0.00 s Inj. Penetration : 40.00 mm Vial Penetration : 30.00 mm CLEANING PARAMETERS Preclean Sample : 0 Preclean Solv.1 : 6 Postclean Solv.1 : 6 Fill Speed Solv.1 : 5.00 uL/s Viscosity Delay Solv.1 : 1.0 s Eject Speed Solv.1 : 50.00 uL/s Preclean Solv.2 : 6 Postclean Solv.2 : 6 Fill Speed Solv.2 : 5.00 uL/s Viscosity Delay Solv.2 : 1.0 s Eject Speed Solv.2 : 50.00 uL/s MS ACQUISITION PARAMETERS General Information ------- ----------- Tune File : atune.u Acquistion Mode : Scan MS Information -- ----------- Solvent Delay : 3.00 min EM Absolute : False EM Offset : 0 Resulting EM Voltage : 1176.5 [Scan Parameters] Low Mass : 41.0 High Mass : 400.0 Threshold : 50 Sample # : 3 A/D Samples 8 [MSZones] MS Quad : 150 C maximum 200 C MS Source : 230 C maximum 250 C END OF MS ACQUISITION PARAMETERS TUNE PARAMETERS for SN: ----------------------------- EMISSION : 34.610 ENERGY : 69.922 REPELLER : 25.935 IONFOCUS : 69.278 ENTRANCE_LE : 9.500 EMVOLTS : 1176.471 AMUGAIN : 2094.000 AMUOFFSET : 134.000 FILAMENT : 1.000 DCPOLARITY : 0.000 ENTLENSOFFS : 17.318 MASSGAIN : -9.000 MASSOFFSET : -10.000 END OF TUNE PARAMETERS ------------------------------------ END OF INSTRUMENT CONTROL PARAMETERS ------------------------------------